level 4 (BSL-4) laboratory A biosafety level (BSL), or pathogen/protection level, is a set of biocontainment precautions required to isolate dangerous biological agents in an enclosed laboratory facility. The levels of containment range from the lowest biosafety level 1 (BSL-1) to the highest at level 4 (BSL-4).
In the United States, the Centers for Disease Control and Prevention (CDC) have specified these levels in a publication referred to as Biosafety in Microbiological and Biomedical Laboratories (BMBL). In the European Union (EU), the same biosafety levels are defined in a directive. In Canada the four levels are known as Containment Levels.
Facilities with these designations are also sometimes given as P1 through P4 (for pathogen or protection level), as in the term P3 laboratory. At the lowest level of biosafety, precautions may consist of regular hand-washing, other hygienic practices, and minimal protective equipment.
At higher biosafety levels, precautions may include airflow systems, multiple containment rooms, sealed containers, positive pressure personnel suits, established protocols for all procedures, extensive personnel training, and high levels of security to control access to the facility. Health Canada reports that world-wide until 1999 there were recorded over 5,000 cases of accidental laboratory infections and 190 deaths.
level 4 (BSL-4) laboratory A biosafety level (BSL), or pathogen/protection level, is a set of biocontainment precautions required to isolate dangerous biological agents in an enclosed laboratory facility. The levels of containment range from the lowest biosafety level 1 (BSL-1) to the highest at level 4 (BSL-4).
In the United States, the Centers for Disease Control and Prevention (CDC) have specified these levels in a publication referred to as Biosafety in Microbiological and Biomedical Laboratories (BMBL). In the European Union (EU), the same biosafety levels are defined in a directive. In Canada the four levels are known as Containment Levels.
Facilities with these designations are also sometimes given as P1 through P4 (for pathogen or protection level), as in the term P3 laboratory. At the lowest level of biosafety, precautions may consist of regular hand-washing, other hygienic practices, and minimal protective equipment.
At higher biosafety levels, precautions may include airflow systems, multiple containment rooms, sealed containers, positive pressure personnel suits, established protocols for all procedures, extensive personnel training, and high levels of security to control access to the facility. Health Canada reports that world-wide until 1999 there were recorded over 5,000 cases of accidental laboratory infections and 190 deaths.
Biosafety level 4
Biosafety level 4 (BSL-4) is the highest level of biosafety precautions, and is appropriate for work with agents that could easily be aerosol-transmitted within the laboratory and cause severe to fatal disease in humans for which there are no available vaccines or treatments. BSL-4 labs are generally set up to be either cabinet laboratories or protective-suit laboratories.
In cabinet laboratories, all work must be done within a class III biosafety cabinet. Materials leaving the cabinet must be decontaminated by passing through an autoclave or a tank of disinfectant. The cabinets themselves are required to have seamless edges to allow easy cleaning.
Also, the cabinet and all materials within must be free of sharp edges to reduce the risk of damage to the gloves. In a protective-suit lab, all work must be done in a class II biosafety cabinet by personnel wearing a positive pressure suit.
To exit a BSL-4 lab, personnel must pass through a chemical shower for decontamination, then a room for removing the positive-pressure suit, followed by a personal shower. Entry into the BSL-4 lab is restricted to trained and authorized individuals, and all persons entering and exiting the lab must be recorded.
As with BSL-3 laboratories, BSL-4 labs must be separated from areas that receive unrestricted traffic. Also, airflow is tightly controlled to ensure that air always flows from "clean" areas of the lab to areas where work with infectious agents is being performed.
The entrance to the BSL-4 lab must also employ airlocks to minimize the possibility that aerosols from the lab could be removed from the lab. All laboratory waste, including filtered air, water, and refuse must also be decontaminated before it can leave the facility.
Biosafety level 4 laboratories are used for diagnostic work and research on easily transmitted pathogens which can cause fatal disease. These include a number of viruses known to cause viral hemorrhagic fever such as Marburg virus, Ebola virus, Lassa virus, and Crimean-Congo hemorrhagic fever. Other pathogens handled at BSL-4 include Hendra virus, Nipah virus, and some flaviviruses.
Also, poorly characterized pathogens which appear closely related to dangerous pathogens are often handled at this level until enough data is obtained either to confirm continued work at this level, or to permit working with them at a lower level. This level is also used for work with smallpox virus, though this work is only performed at the Centers for Disease Control and Prevention in Atlanta, United States, and the State Research Center of Virology and Biotechnology in Koltsovo, Russia.
File:NIAID Integrated Research Facility - Positive Pressure Personnel Suit Inspection.jpg|Regular inspection of positive-pressure suits to locate any leaks File:NIAID Integrated Research Facility - SPECT.jpg|alt=SPECT machine at BSL-4 imaging facility that separates subjects with pathogens from the machines.|SPECT machine at BSL-4 imaging facility that separates subjects with pathogens from the machines File:NIAID Integrated Research Facility - MRI.jpg|The circular containment tube separates the patient table in the "hot" zone (pathogen present) from the "cold" zone around this MRI machine. File:NIAID Integrated Research Facility - Air Pressure Resistant (APR) door.jpg|Air pressure resistant (APR) door to separate the hot and cold zones File:NIAID Integrated Research Facility - Positive Pressure Personnel Suit.jpg|Working inside a BSL-4 lab with air hoses providing positive air pressure File:NIAID Integrated Research Facility - Class III Biosafety Cabinet.jpg|Inside a Class III biological safety cabinet with an aerosol control platform File:Effluent decontamination system.jpg|Effluent decontamination system of a BSL-4 lab of NIAID
Safety concerns
A North Carolina Mosquito & Vector Control Association (NCMVCA) study highlighted safety concerns. In the United States, laboratories can be funded by federal, state, private, non-profit, or academically. The last accounts for 72% of the funding.
High-containment labs that are registered with the Centers for Disease Control and Prevention (CDC) and the U.S. Department of Agriculture's (USDA) Select Agent Program must adhere to Department of Defense standards. Since BSL3 and 4 laboratories in the United States are regulated by either the CDC or USDA or another federal agency (depending on the pathogens they handle), no single federal agency is responsible for regulating or tracking the number of these labs.
U.S. high-containment laboratories that handle pathogens which are declared as "select agents" must be inspected periodically by the CDC or USDA, adhere to certain standards, and maintain ongoing education on biosecurity and biosafety policies as mandated by law.