Cleanroom
≈ 7 min read
Names in other languages
No names in other languages are held for this entry yet.
Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.
(red-green) lighting is necessary for photolithography, to prevent unwanted exposure of photoresist to light of shorter wavelengths. manufacturing with fan filter units installed in the ceiling grid A cleanroom or clean room is an engineered space that maintains a very low concentration of airborne particulates. It is well-isolated, well-controlled from contamination, and actively cleansed. Such rooms are commonly needed for scientific research and in industrial production for all nanoscale processes, such as semiconductor device manufacturing. A cleanroom is designed to keep everything from dust to airborne organisms or vaporised particles away from it, and so from whatever material is being handled inside it. A cleanroom can also prevent the escape of materials. This is often the primary aim in hazardous biology, nuclear work, pharmaceutics, and virology. Cleanrooms typically come with a cleanliness level quantified by the number of particles per cubic meter at a predetermined molecule measure. The ambient outdoor air in a typical urban area contains 35,000,000 particles for each cubic meter in the size range 0.5 μm and bigger, equivalent to an ISO 9 certified cleanroom. By comparison, an ISO 14644-1 level 1 certified cleanroom permits no particles in that size range, and just 12 particles for each cubic meter of 0.3 μm and smaller.…
(red-green) lighting is necessary for photolithography, to prevent unwanted exposure of photoresist to light of shorter wavelengths. manufacturing with fan filter units installed in the ceiling grid A cleanroom or clean room is an engineered space that maintains a very low concentration of airborne particulates. It is well-isolated, well-controlled from contamination, and actively cleansed. Such rooms are commonly needed for scientific research and in industrial production for all nanoscale processes, such as semiconductor device manufacturing. A cleanroom is designed to keep everything from dust to airborne organisms or vaporised particles away from it, and so from whatever material is being handled inside it. A cleanroom can also prevent the escape of materials. This is often the primary aim in hazardous biology, nuclear work, pharmaceutics, and virology. Cleanrooms typically come with a cleanliness level quantified by the number of particles per cubic meter at a predetermined molecule measure. The ambient outdoor air in a typical urban area contains 35,000,000 particles for each cubic meter in the size range 0.5 μm and bigger, equivalent to an ISO 9 certified cleanroom. By comparison, an ISO 14644-1 level 1 certified cleanroom permits no particles in that size range, and just 12 particles for each cubic meter of 0.3 μm and smaller. Semiconductor facilities often get by with level 7 or 5, while level 1 facilities are exceedingly rare.
History
The modern cleanroom was invented by American physicist Willis Whitfield. As an employee of the Sandia National Laboratories, Whitfield created the initial plans for the cleanroom in 1960. Prior to Whitfield's invention, earlier cleanrooms often had problems with particles and unpredictable airflows. Whitfield designed his cleanroom with a constant, highly filtered airflow to flush out impurities. Within a few years of its invention in the 1960s, Whitfield's modern cleanroom had generated more than US$50 billion in sales worldwide (approximately $ billion today). By mid-1963, more than 200 U.S. industrial plants had such specially constructed facilities—then using the terminology "white rooms", "clean rooms", or "dust-free rooms"—including the Radio Corporation of America, McDonnell Aircraft, Hughes Aircraft, Sperry Rand, Sylvania Electric, Western Electric, Boeing, and North American Aviation. RCA began such a conversion of part of its Cambridge, Ohio facilities in February 1961. Totalling 70,000 square feet, it was used to prepare control equipment for the Minuteman ICBM missiles. The majority of the integrated circuit manufacturing facilities in Silicon Valley were made by three companies: MicroAire, PureAire, and Key Plastics. These competitors made laminar flow units, glove boxes, cleanrooms and air showers, along with the chemical tanks and benches used in the "wet process" building of integrated circuits. These three companies were the pioneers of the use of Teflon for airguns, chemical pumps, scrubbers, water guns, and other devices needed for the production of integrated circuits. William (Bill) C. McElroy Jr. worked as an engineering manager, drafting room supervisor, QA/QC, and designer for all three companies, and his designs added 45 original patents to the technology of the time. McElroy also wrote a four-page article for MicroContamination Journal, wet processing training manuals, and equipment manuals for wet processing and cleanrooms.
Personnel contamination of cleanrooms
The greatest threat to cleanroom contamination comes from the users themselves. In the healthcare and pharmaceutical sectors, control of microorganisms is important, especially microorganisms likely to be deposited into the air stream from skin shedding. Studying cleanroom microflora is of importance for microbiologists and quality control personnel to assess changes in trends. Shifts in the types of microflora may indicate deviations from the "norm" such as resistant strains or problems with cleaning practices. In assessing cleanroom microorganisms, the typical flora are primarily those associated with human skin (Gram-positive cocci), although microorganisms from other sources such as the environment (Gram-positive rods) and water (Gram-negative rods) are also detected, although in lower number. Common bacterial genera include Micrococcus, Staphylococcus, Corynebacterium, and Bacillus, and fungal genera include Aspergillus and Penicillium. \text{C}_\text{N} = 10^{\text{N}} \left(\frac{0.1}{\text{D}}\right)^{2.08} Where \text{C}_\text{N} is the maximum concentration of particles in a volume of 1m ^3 of airborne particles that are equal to, or larger, than the considered particle size which is rounded to the nearest whole number, using no more than three significant figures, \text{N} is the ISO class number, \text{D} is the size of the particle in \mu m and 0.1 is a constant expressed in \mu m. The result for standard particle sizes is expressed in the following table.
US FED STD 209E
US FED-STD-209E was a United States federal standard. It was officially cancelled by the General Services Administration on November 29, 2001, but is still widely used. Current regulating bodies include ISO, USP 800, US FED STD 209E (previous standard, still used). Drug Quality and Security Act (DQSA) created in Nov. 2013 in response to drug compounding deaths and serious adverse events. The Federal Food, Drug, and Cosmetic Act (FD&C Act) created specific guidelines and policies for human compounding. 503A addresses compounding by state or federally licensed facility by licensed personnel (pharmacist/ physicians) 503B pertaining to outsourcing facilities need direct supervision from a licensed pharmacist and do not need to be a licensed pharmacy. Facility is licensed through the Food and Drug Administration (FDA)
EU GMP classification
EU GMP guidelines are more stringent than others, requiring cleanrooms to meet particle counts at operation (during manufacturing process) and at rest (when manufacturing process is not carried out, but room AHU is on).
BS 5295
BS 5295 is a British Standard. BS 5295 Class 1 also requires that the greatest particle present in any sample can not exceed 5 μm. BS 5295 has been superseded, withdrawn since the year 2007 and replaced with "BS EN ISO 14644-6:2007".
See also
Air quality index Contamination control Data recovery lab Indoor air quality Microfiltration Pressure room (disambiguation) Pneumatic filter Secure environment
References
External links
Cleanroom Wiki--The Global Society For Contamination Control (GSFCC)
Text derived from https://en.wikipedia.org/w/index.php?oldid=1378384206 under CC BY-SA 4.0; modifications have been made.