Digital ion trap

The digital ion trap (DIT) is an quadrupole ion trap driven by digital signals, typically in a rectangular waveform, generated by switching rapidly between discrete DC voltage levels.

Uses your system voice; voice and rate in settings.

Snapshot

≈ 8 min read

Names in other languages

No names in other languages are held for this entry yet.

Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.

The digital ion trap (DIT) is an quadrupole ion trap driven by digital signals, typically in a rectangular waveform, generated by switching rapidly between discrete DC voltage levels. The digital ion trap has been mainly developed as a mass analyzer. a_z = -\frac {8eU} {m r_0^2 \Omega^2} \qquad\qquad (1) \! and q_z = \frac {4eV} {m r_0^2 \Omega^2} . \qquad\qquad (2) \!

Here, Ω =2πf is the angular frequency of the digital waveform. Similar definitions of the a , q for the 2D (linear) ion trap were also given in literature. There are at least two postulates about the nature of the DC component.

The first, of which has been attributed to Ding, assumes for the DIT that the DC component, U depends on not only the mid-level of the AC voltages, V 1 and V 2 , but also the duty cycle, d of the waveform: U = dV_1 + (1-d) V_2 Whereas, the second but more general postulate assumes that there is no DC component unless there is an explicit DC voltage offset added to the waveforms. The latter interpretation is explained by the change to the stability diagram that results when the duty cycle moves away from d = 0.5.

When this happens the range of stable q and a values for both quadrupole axes change. These changes cause the motion of ions to be more displaced along one axis compared to the other. This, consequently is the effect of the DC bias.

It is important to accurately know the stability of ions inside the DIT. For example, different waveform duty cycles result in a different stability boundary. For the case of a square wave, where d = 0.5, the boundary of the first stability region crosses the q_z axis at approximately 0.712, which is less than 0.908, the boundary value q_z for a sinusoidal waveform.

The stability of ion motion in a digitally driven quadrupole can be calculated from the analytical matrix solutions of Hill's equation: \mathbf V(f_n, \tau_n)=\begin{bmatrix} cos(\tau_n\sqrt{f_n)} & 1/\sqrt{f_n} sin(\tau_n\sqrt{f_n}\\ -\sqrt{f_n}sin(\tau_n\sqrt{f_n}) & cos(\tau_n\sqrt{f_n}) \end{bmatrix} \qquad f_n >0 \qquad\qquad\qquad (3a) \mathbf V(f_n, \tau_n)=\begin{bmatrix} cosh(\tau_n\sqrt{-f_n)} & 1/\sqrt{-f_n} sinh(\tau_n\sqrt{-f_n}\\ \sqrt{-f_n}sinh(\tau_n\sqrt{-f_n}) & cosh(\tau_n\sqrt{-f_n}) \end{bmatrix} \qquad f_n The analytical solutions apply to any periodic function so long as each period, T can be represented as a series of n constant potential steps T = \sum^n_1 t_n . Each constant potential step is represented in dimensionless Mathieu space by the waveform potential parameter f = a\pm2q , where q and a were defined earlier by (1) and (2). The value \tau_n = t_n\pi in (3) is the temporal width of the constant potential step.

In a digital system that is operated without a physical DC offset the waveform potential reduces to the value \pm 2q . The sign of the parameter will depend on the sign of the constant potential at each step, and the appropriate matrix will depend on the sign of the parameter. Because a digital waveform may be approximated as existing in only high and low states (potential sign), the stability of ions, as demonstrated by Brabeck, may be determined in as few as two or three constant potential steps.

In the simple but frequent case that a full cycle of a digital waveform can be represented by two constant potential steps, the matrix representing the first potential step would be multiplied onto the matrix representing the second potential step. In the general case, the final matrix of a waveform cycle defined by n constant potential steps is: \mathbf M=\mathbf V(f_1,\tau_1)\times \mathbf V(f_2,\tau_2)...\times \mathbf V(f_n, \tau_n) \qquad\qquad\qquad(4) The matrix (4) is often referred to as the transfer matrix.

It is used to evaluate whether an ion will have stable motion. If the absolute value of the trace of this matrix is less than 2 the ion is said to have stable motion. Stable motion simply means that the secular oscillation of the ion has a maximum displacement.

When the absolute value of the trace is greater than 2 ion motion is not stable and the displacement of the ion increases with each secular oscillation. Ion trajectories in a linear or 3D DIT as well as in a digital mass filter, may also be calculated using a similar procedure.

Unlike stability calculation it is advantageous for the purpose of resolution and accuracy to represent each period of the waveform with an adequate number of constant voltage steps. The trajectory for the constant potential step, k , for example, is calculated by multiplication of the appropriate matrix (3) for that step onto the trajectory vector of the step, k-1 : \binom {u_k}{\dot u_k}=\mathbf V(f_k, \tau_k)\binom{u_{k-1}}{\dot u_{k-1}} \qquad \qquad (5) A stability diagram may be generated by calculating the matrix trace for each axis over a defined range of q and a values.

The stability diagram of a square wave is very similar to that of the traditional harmonic quadrupole field. Having the additional parameter d in the waveform, the digital ion trap can perform certain experiments which are not available in the conventional harmonic wave RF ion trap.

One example is the digital asymmetric wave isolation which is the method of using a d value around 0.6 to narrow the mass range to isolate a precursor ion. The DIT is a versatile instrument because it is capable of operating at constant AC voltage without a DC offset for any conceivable duty cycle and frequency. The dynamic frequency does not impose a limit on the mass range.

The Mathieu space stability diagram of the linear and 3D DIT change with duty cycle. When a = 0 there will be a finite range of stable q values for each quadrupole axis that will depend on the duty cycle. Fig 3 (a) shows a Mathieu space stability diagram for the duty cycle d = 0.50 of a linear DIT.

The horizontal line indicates where the parameter a = 0. The range of completely stable q values appear where this line passes through the green colored region; it ranges from q = 0 to roughly q = 0.7125. The blue colored areas in the figure depict stability along the x-axis only.

Yellow colored regions depict stability along the y-axis only. When the duty cycle is increased to d = 0.60 the range of completely stable q values decreases (see Fig 3 (b)) as indicated by the reduction of green that the horizontal line intersects.

In this representation the total range of stable q values along the x-axis, that is defined by the intersection of the line through the blue and green regions, is greater than the total range of stable q values along the y-axis that is defined by the intersection of the line through the yellow and green regions. In Fig 3 (b) the overall stability of the linear DIT in the y direction is smaller than in the x direction.

If the frequency of the linear DIT is decreased to cause a particular ion to have a q value that corresponds to right hand side boundary of the completely stable green region, then it will excite and ultimately eject in the y direction. This is the fundamental mechanism that allows control over the direction of ion excitation in a linear DIT without resonant excitation.

The DIT and other forms of digital mass analyzers scan ions by scanning the frequency of the drive waveform. The AC voltage is typically fixed during the scan. Digital devices use a duty cycle which allows them to operate completely independent of a DC voltage and without resonant excitation.

When the DC voltage is zero the parameter a is also zero. Consequently, ion stability will depend on q. With these considerations it was possible to design a new type of stability diagram that is more suitable for planning and performing experiment.

In 2014 Brabeck and Reilly created a stability diagram that maps the range of stable mass-to-charge ratios, m/z to the corresponding range of drive frequencies based on several user inputs. For a particular duty cycle, the operator can quickly reference the range of stable masses at each frequency of a scan.

Fig 4 (a) and (b) shows the frequency-m/z stability diagram for a linear DIT with a duty cycle of d = 0.50 and d= 0.60 respectively. \omega_z = \frac {1} {2} \beta_z\Omega \qquad\qquad (6) \! For digital driving signal, Ding derived the expression of the secular frequency using matrix transform theory. \omega_z = \frac {\Omega} {2\pi} arccos \frac {\phi_{11} + \phi_{22} } {2} Where : \phi_{11}, \phi_{22} are two diagonal elements of the transform matrix of ion motion.

For a DC free square wave ( a =0 ) the transform matrix may be expressed using the stability parameter q , thus: \beta_z = \frac {1} {\pi} arccos[cos (\pi{\sqrt{q_z/2}})cosh ( \pi {\sqrt{q_z/2}}) \qquad\qquad (7) \! The formula (6) and (7) give a direct relation between the secular frequency and the digital drive waveform parameters (frequency and amplitude), without using the iterative process that is needed for a sinusoidal driven quadrupole ion trap.

Normally the depth of the 'effective potential' well, or the pseudopotential well, is used to estimate the maximum kinetic energy of ions that remain trapped. For DIT, this was also derived using Dehmelt approximation: D_z = \frac {{\pi}^2}{48} q_z V \approx 0.206 q_z V [eV]

Text derived from https://en.wikipedia.org/w/index.php?oldid=1325837792 under CC BY-SA 4.0; modifications have been made.