Digital microfluidics
fluid handling technique
≈ 6 min read
Names in other languages
- العربيةعلم الموائع الدقيق الرقمي
- فارسیریزسیالشناسی دیجیتالی
- EnglishDigital microfluidics
- 日本語デジタルマイクロ流体力学
Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.
Digital microfluidics (DMF) is a platform for lab-on-a-chip systems that is based upon the manipulation of microdroplets. Droplets are dispensed, moved, stored, mixed, reacted, or analyzed on a platform with a set of insulated electrodes.
Digital microfluidics can be used together with analytical analysis procedures such as mass spectrometry, colorimetry, electrochemical, and electrochemiluminescense. [[File:DMF open and closed system.png|alt=|right|frameless|577x577px| Aqueous droplet sitting on top of an open and closed digital microfluidic system with a cross section view.
This shows the movement of the droplet once an electrode is activated. Device design can be manipulated to fit user's needs (modified electrodes, electrode pattern, materials used, etc.).[3][4 ] A common substrate used in this type of system is glass. Depending if the system is open or closed, there would be either one or two layers of glass.
The bottom layer of the device contains a patterned array of individually controllable electrodes. When looking at a closed system, there is usually a continuous ground electrode found through the top layer made usually of indium tin oxide (ITO).
The dielectric layer is found around the electrodes in the bottom layer of the device and is important for building up charges and electrical field gradients on the device. A hydrophobic layer is applied to the top layer of the system to decrease the surface energy where the droplet will actually be in contact with.
The applied voltage activates the electrodes and allows changes in the wettability of droplet on the device's surface. In order to move a droplet, a control voltage is applied to an electrode adjacent to the droplet, and at the same time, the electrode just under the droplet is deactivated.
By varying the electric potential along a linear array of electrodes, electrowetting can be used to move droplets along this line of electrodes. Modifications to this foundation can also be fabricated into the basic design structure.
One example of this is the addition of electrochemiluminescence detectors within the indium tin oxide layer (the ground electrode in a closed system) which aid in the detection of luminophores in droplets. In general, different materials may also be used to replace basic components of a DMF system such as the use of PDMS instead of glass for the substrate.
Liquid materials can be added, such as oil or another substance, to a closed system to prevent evaporation of materials and decrease surface contamination. Also, DMF systems can be compatible with ionic liquid droplets with the use of an oil in a closed device or with the use of a catena (a suspended wire) over an open DMF device. Digital microfluidics can be light-activated.
Optoelectrowetting can be used to transport sessile droplets around a surface containing patterned photoconductors. The photoelectrowetting effect can also be used to achieve droplet transport on a silicon wafer without the necessity of patterned electrodes. \cos(\theta)=\cos(\theta{_0})+\frac{\varepsilon{_0}\varepsilon{_r}V^2}{{2\gamma}d} where \theta is the contact angle with an applied voltage V ; \theta{_0} is the contact angle with no voltage; \varepsilon{_r} is the relative permittivity of the dielectric; \varepsilon{_0} is the permittivity of free space; \gamma is the liquid/filler media surface tension; d is the dielectric thickness.
In some cases, the hydrophobicity of a substrate can be controlled by using electrical fields. This refers to the phenomenon Electrowetting On Dielectric (EWOD).[3][4] For example, when no electric field is applied to an electrode, the surface will remain hydrophobic and a liquid droplet will form a more spherical droplet with a greater contact angle.
When an electric field is applied, a polarized hydrophilic surface is created. The water droplet then becomes flattened and the contact angle decreases. By controlling the localization of this polarization, we can create an interfacial tension gradient that allows controlled displacement of the droplet across the surface of the DMF device.
A droplet can be split by charging two electrodes on opposite sides of a droplet on an uncharged electrode. In the same way a droplet on an uncharged electrode will move towards an adjacent, charged electrode, this droplet will move towards both active electrodes. Liquid moves to either side, which causes the middle of the droplet to neck.
For a droplet of the same size as the electrodes, splitting will occur approximately when R_{neck}/R_{end}=-1 , as the neck will be at its thinnest. R_{neck} is the radius of curvature of the menisci at the neck, which is negative for a concave curve, and R_{end} is the radius of curvature of the menisci at the elongated ends of the droplet.
This process is simple and consistently results in two droplets of equal volume. The conventional method of splitting an existing droplet by simply turning the splitting electrodes on and off produces new droplets of relatively equal volume. However, the new droplets formed by the conventional method show considerable difference in volume.
This difference is caused by local perturbations due to the rapid mass transport. Even though the difference is negligible in some applications, it can still pose a problem in applications that are highly sensitive to variations in volume, such as immunoassays and DNA amplification.
To overcome the limitation of the conventional method, an existing droplet can be split by gradually changing the potential of the electrodes at the splitting region instead of simply switching them on and off. Using this method, a noticeable improvement in droplet volume variation, from around 10% variation in volume to less than 1% variation in volume, has been reported.
Nuclear magnetic resonance spectroscopy
Nuclear magnetic resonance (NMR) spectroscopy can be used in conjunction with digital microfluidics (DMF) through the use of NMR microcoils, which are electromagnetic conducting coils that are less than 1 mm in size. Due to their size, these microcoils have several limitations, directly influencing the sensitivity of the machinery they operate within.
Microchannel/microcoil interfaces, previous to digital microfluidics, had several drawbacks such as in that many created large amounts of solvent waste and were easily contaminated. In this way, the use of digital microfluidics and its capability to manipulate singlet droplets is promising.
The interface between digital microfluidics and NMR relaxometry has led to the creation of systems such as those used to detect and quantify the concentrations of specific molecules on microscales with some such systems using two step processes in which DMF devices guide droplets to the NMR detection site. Introductory systems of high-field NMR and 2D NMR in conjunction with microfluidics have also been developed.
These systems use single plate DMF devices with NMR microcoils in place of the second plate. Recently, further modified version of this interface included pulsed field gradients (PFG) units that enabled this platform to perform more sophisticated NMR measurements (e.g. NMR diffusometry, gradients encoded pulse measurements).
This system has been successfully applied into monitoring rapid organic reactions.