Precession electron diffraction (PED) is a specialized method to collect electron diffraction patterns in a transmission electron microscope (TEM). By rotating (precessing) a tilted incident electron beam around the central axis of the microscope, a PED pattern is formed by integration over a collection of diffraction conditions. This produces a quasi-kinematical diffraction pattern that is more suitable as input into direct methods algorithms to determine the crystal structure of the sample.

Precession electron diffraction (PED) is a specialized method to collect electron diffraction patterns in a transmission electron microscope (TEM). By rotating (precessing) a tilted incident electron beam around the central axis of the microscope, a PED pattern is formed by integration over a collection of diffraction conditions. This produces a quasi-kinematical diffraction pattern that is more suitable as input into direct methods algorithms to determine the crystal structure of the sample. d \propto 4 C_s \phi^2 \alpha Thus, if the specimen of interest is quite small, the maximum precession angle will be restrained. This is most significant for conditions of convergent beam illumination. 50 nm is a general lower limit on probe size for standard TEMs operating at high precession angles (>30 mrad), but can be surpassed in C s corrected instruments. In principle the minimum precessed probe can reach approximately the full-width-half-max (FWHM) of the converged un-precessed probe in any instrument, however in practice the effective precessed probe is typically ~10-50x larger due to uncontrolled aberrations present at high angles of tilt. For example, a 2 nm precessed probe with >40 mrad precession angle was demonstrated in an aberration-corrected Nion UltraSTEM with native sub-Å probe (aberrations corrected to ~35 mrad half-angle). If the precession angle is made too large, further complications due to the overlap of the ZOLZ and HOLZ reflections in the projected pattern can occur. This complicates the indexing of the diffraction pattern and can corrupt the measured intensities of reflections near the overlap region, thereby reducing the effectiveness of the collected pattern for direct methods calculations. \begin{align} I_\mathbf{g}^{kinematical}=|F_\mathbf{g} |^2 \end{align} This relationship is generally far from accurate for experimental dynamical electron diffraction and when many reflections have a large excitation error. First, a Lorentz correction analogous to that used in x-ray diffraction can be applied to account for the fact that reflections are infrequently exactly at the Bragg condition over the course of a PED measurement. This geometrical correction factor can be shown to assume the approximate form: I_\mathbf{g}^{kinematical} \propto I_\mathbf{g}^{experimental} \cdot g\sqrt{1-\frac{g}{2R_o}} where g is the reciprocal space magnitude of the reflection in question and R o is the radius of the Laue circle, usually taken to be equal to φ. While this correction accounts for the integration over the excitation error, it takes no account for the dynamical effects that are ever-present in electron diffraction. This has been accounted for using a two-beam correction following the form of the Blackman correction originally developed for powder x-ray diffraction. Combining this with the aforementioned Lorentz correction yields: I_\mathbf{g}^{kinematical} \propto I_\mathbf{g}^{experimental} \cdot g\sqrt{1-\frac{g}{2R_o}} \cdot \int\limits_0^{A_\mathbf{g}}J_0(2x)\, dx where A_\mathbf{g} = \frac{2 \pi t F_\mathbf{g}}{k} , t is the sample thickness, and k is the wave-vector of the electron beam. J_0 is the Bessel function of zeroeth order. This form seeks to correct for both geometric and dynamical effects, but is still only an approximation that often fails to significantly improve the kinematic quality of the diffraction pattern (sometimes even worsening it). More complete and accurate treatments of these theoretical correction factors have been shown to adjust measured intensities into better agreement with kinematical patterns. For details, see Chapter 4 of reference. Only by considering the full dynamical model through multislice calculations can the diffraction patterns generated by PED be simulated. However, this requires the crystal potential to be known, and thus is most valuable in refining the crystal potentials suggested through direct methods approaches. The theory of precession electron diffraction is still an active area of research, and efforts to improve on the ability to correct measured intensities without a priori knowledge are ongoing.

Text derived from https://en.wikipedia.org/w/index.php?oldid=1377335337 under CC BY-SA 4.0; modifications have been made.

Sources cited in this revision

7 named source(s) extracted from the cited revision. A source with no author is shown by its publication; nothing is invented.