Probe electrospray ionization

Probe electrospray ionization (PESI) is an electrospray-based ambient ionization technique which is coupled with mass spectrometry for sample analysis.

Uses your system voice; voice and rate in settings.

≈ 4 min read

Cite, export and learn8 citation formats · 8 exports · lesson and quiz

Names in other languages

No names in other languages are held for this entry yet.

Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.

Probe electrospray ionization (PESI) is an electrospray-based ambient ionization technique which is coupled with mass spectrometry for sample analysis. Unlike traditional mass spectrometry ion sources which must be maintained in a vacuum, ambient ionization techniques permit sample ionization under ambient conditions, allowing for the high-throughput analysis of samples in their native state, often with minimal or no sample pre-treatment.

The PESI ion source simply consists of a needle to which a high voltage is applied following sample pick-up, initiating electrospray directly from the solid needle.

History

Probe electrospray ionization is an ambient ionization mass spectrometry technique developed by Kenzo Hiraoka et al. at the University of Yamanashi, Japan. The technique was developed to address some of the issues associated with traditional electrospray ionization (ESI), including clogging of the capillary and contamination, whilst providing a means of rapid and direct sample analysis.

Since its initial conception, various modified forms of the PESI ion source have been developed, and the PESI-MS system has been commercialized by instrument manufacturing company Shimadzu.

Principle of operation

The PESI ion source consists of a solid needle or wire which acts as both the sampling probe and electrospray emitter. The needle is moved up and down along a vertical axis, a process which can be either automated or manual. When the needle is lowered to the sampling stage, the tip of the needle briefly touches the surface of a typical liquid sample.

During this stage, the needle is held at ground potential. The needle is then raised to be level with the mass spectrometer inlet where a high voltage of 2–3 kV is applied. Electrospray is induced at the tip of the needle, producing analyte ions which are drawn into the mass spectrometer for analysis.

The mechanism by which ions are formed is believed to be identical to traditional electrospray ionization. As a result, in positive ion mode analytes are often observed as the protonated, sodiated and potentiated ions, depending on the sample and analyte type.

Although the amount of sample picked up by the needle is largely dependent on sample viscosity, it has been estimated that just a few picolitres of the sample solution are typically used. Because of this, the technique can be applied to small sample sizes, particularly ideal when limited sample amounts are available.

As such a small sample amount is picked up and completely exhausted during the ionization process, issues of contamination are severely reduced. Furthermore, the process of sampling and ionization takes just a few seconds, so PESI-MS is suitable for high-throughput analysis.

Sequential ionization

A phenomenon observed with probe electrospray ionization is the sequential and exhaustive ionization of analytes with different surface activities. During the development of PESI, it was discovered that analytes could be sequentially ionized throughout the electrospray, thus enabling a temporal separation of components within a sample.

In normal ESI, the sample solution is typically continuously supplied through a capillary and the charged droplets contain all sample components, with more surface-active analytes being constantly preferentially ionized. In PESI, surface-active analytes are also preferentially ionized.

However, as a finite droplet exists on the tip of the needle, following the depletion of surface-active analytes, the remaining components in the droplet can then be ionized and observed. This can result in the production of distinctively different mass spectra from a single sample over the application of the high voltage for just a few seconds.

This effect offers a particular advantage in the analysis of analytes suffering from ion suppression effects. The presence of surface-active analytes or charged solvent additives can result in the suppressed ionization of analytes of interest, resulting in low sensitivity or the complete absence of the analyte.

The effects of ion suppression can be minimized by reducing the complexity of the sample, for instance through sample extraction techniques such as solid phase extraction, or by separation of analytes of interest using chromatographic separation. However, these sample preparation steps can be laborious, time-consuming and expensive.

PESI enables a reduction in ion suppression without the need for sample pre-treatment. By separating the ionization of different analytes, components causing ion suppression can be exhausted before enabling the ionization of components of interest.

This has been demonstrated in a number of scenarios, including in the analysis of raw urine, with concentrated components such as creatinine ionization initially, followed by the appearance of previously undetected metabolites.

Text derived from https://en.wikipedia.org/w/index.php?oldid=1326188517 under CC BY-SA 4.0; modifications have been made.

Words in each sectionSequential ionizationSequential ionizationSequential ionization: 321 words321Principle of operationPrinciple of operationPrinciple of operation: 252 words252HistoryHistoryHistory: 82 words82
Words in each section. Measured from the stored article body; sections below the minimum are not plotted.

Provenance

snapshot enwiki-001326188517 · captured 2025-12-07 · ledger · version history

Cite this page as: encyclopegia.org, “Probe electrospray ionization”, snapshot enwiki-001326188517. https://encyclopegia.org/a/probe-electrospray-ionization

Sources cited in this revision

1 named source(s) extracted from the cited revision. A source with no author is shown by its publication; nothing is invented.