Scanning electron microscope
≈ 22 min read
Names in other languages
No names in other languages are held for this entry yet.
Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.
s taken on a SEM shows the characteristic depth of field of SEM micrographs first SEM A scanning electron microscope (SEM) is a type of electron microscope that produces images of a sample by scanning the surface with a focused beam of electrons. The electrons interact with atoms in the sample, producing various signals that contain information about the surface topography and composition. The electron beam is scanned in a raster scan pattern, and the position of the beam is combined with the intensity of the detected signal to produce an image. In the most common SEM mode, secondary electrons emitted by atoms excited by the electron beam are detected using a secondary electron detector (Everhart–Thornley detector). The number of secondary electrons that can be detected, and thus the signal intensity, depends, among other things, on specimen topography. Some SEMs can achieve resolutions better than 1 nanometer. Specimens are observed in high vacuum in a conventional SEM, or in low vacuum or wet conditions in a variable pressure or environmental SEM, and at a wide range of cryogenic or elevated temperatures with specialized instruments.
s taken on a SEM shows the characteristic depth of field of SEM micrographs first SEM A scanning electron microscope (SEM) is a type of electron microscope that produces images of a sample by scanning the surface with a focused beam of electrons. The electrons interact with atoms in the sample, producing various signals that contain information about the surface topography and composition. The electron beam is scanned in a raster scan pattern, and the position of the beam is combined with the intensity of the detected signal to produce an image. In the most common SEM mode, secondary electrons emitted by atoms excited by the electron beam are detected using a secondary electron detector (Everhart–Thornley detector). The number of secondary electrons that can be detected, and thus the signal intensity, depends, among other things, on specimen topography. Some SEMs can achieve resolutions better than 1 nanometer. Specimens are observed in high vacuum in a conventional SEM, or in low vacuum or wet conditions in a variable pressure or environmental SEM, and at a wide range of cryogenic or elevated temperatures with specialized instruments.
History
An account of the early history of scanning electron microscopy has been presented by McMullan. Although Max Knoll produced a photo with a 50 mm object-field-width showing channeling contrast by the use of an electron beam scanner, it was Manfred von Ardenne who in 1937 invented a microscope with high resolution by scanning a very small raster with a demagnified and finely focused electron beam. In the same year, Cecil E. Hall also completed the construction of the first emission microscope in North America, just two years after being tasked by his supervisor, E. F. Burton at the University of Toronto. Ardenne applied scanning of the electron beam in an attempt to surpass the resolution of the transmission electron microscope (TEM), as well as to mitigate substantial problems with chromatic aberration inherent to real imaging in the TEM. He further discussed the various detection modes, possibilities and theory of SEM, together with the construction of the first high resolution SEM. Further work was reported by Zworykin's group, followed by the Cambridge groups in the 1950s and early 1960s headed by Charles Oatley, all of which finally led to the marketing of the first commercial instrument by Cambridge Scientific Instrument Company as the "Stereoscan" in 1965, which was delivered to DuPont.
Principles and capacities
The signals used by an SEM to make an image result from interactions between the electron beam and atoms at various depths within the sample. Various types of signals are produced including secondary electrons (SE), reflected or back-scattered electrons (BSE), characteristic X-rays and light (cathodoluminescence) (CL), absorbed current (specimen current) and transmitted electrons. Secondary electron imaging and back-scattered electron detectors are standard approaches in an SEM , but additional detectors may be used to capture additional signals. For instance, emitted X-rays can be detected by energy dispersive X-ray spectrometry. Secondary electrons have very low energies on the order of 50 eV, which limits their mean free path in solid matter. Consequently, SEs can only escape from the top few nanometers of the surface of a sample. The signal from secondary electrons tends to be highly localized at the point of impact of the primary electron beam, making it possible to collect images of the sample surface with a resolution of below 1 nm. Back-scattered electrons (BSE) are beam electrons that are reflected from the sample by elastic scattering. Since they have much higher energy than SEs, they emerge from deeper locations within the specimen and, consequently, the resolution of BSE images is less than SE images. However, BSE are often used in analytical SEM, along with the spectra made from the characteristic X-rays, because the intensity of the BSE signal is strongly related to the atomic number (Z) of the specimen. BSE images can provide information about the distribution, but not the identity, of different elements in the sample. In samples predominantly composed of light elements, such as biological specimens, BSE imaging can image colloidal gold immuno-labels of 5 or 10 nm diameter, which would otherwise be difficult or impossible to detect in secondary electron images. Characteristic X-rays are emitted when the electron beam removes an inner shell electron from the sample, causing a higher-energy electron to fill the shell and release energy. The energy or wavelength of these characteristic X-rays can be measured by Energy-dispersive X-ray spectroscopy or Wavelength-dispersive X-ray spectroscopy and used to identify and measure the abundance of elements in the sample and map their distribution. Due to the very narrow electron beam, SEM micrographs have a large depth of field yielding a characteristic three-dimensional appearance useful for understanding the surface structure of a sample. This is exemplified by the micrograph of pollen shown above. A wide range of magnifications is possible, from about 10 times (about equivalent to that of a powerful hand-lens) to more than 500,000 times, about 250 times the magnification limit of the best light microscopes.
Sample preparation
in gold, having been prepared for viewing with an SEM . No conductive coating was applied: such a coating would alter this fragile specimen. SEM samples have to be small enough to fit on the sample stage, secured with a conductive paint. The relatively low vacuum (0.076 to 20 torr) of Variable-Pressure and Environmental SEM obviates the need for cleaning, even those with water or oily films, allowing observation in a natural state. Preparation is identical to the steps used in light microscopy polished samples. SEM techniques can be used to examine integrated circuit condition and attachment points. Foreshortening is achieved by tilting any specimen. Sputter coating is used to apply a thin coating of metal, or metal alloy, to nonconductive specimens. Conductive materials in current use for specimen coating include gold, gold/palladium alloy, platinum, iridium, tungsten, chromium, osmium. . An alternative to coating for some biological samples is to increase the bulk conductivity of the material by impregnation with osmium using variants of the OTO staining method (O-osmium tetroxide, T-thiocarbohydrazide, O-osmium).
Biological samples
Since the SEM specimen chamber is under high vacuum, a SEM specimen must be completely dry or cryogenically cooled. Hard, dry materials such as wood, bone, feathers, dried insects, or shells (including egg shells ) can be examined with little further treatment, but living cells and tissues and whole, soft-bodied organisms require chemical fixation to preserve and stabilize their structure. Fixation is usually performed by incubation in a solution of a buffered chemical fixative, such as glutaraldehyde, sometimes in combination with formaldehyde and other fixatives, and optionally followed by postfixation with osmium tetroxide. The fixed tissue is then dehydrated. Because air-drying causes collapse and shrinkage, this is commonly achieved by replacement of water in the cells with organic solvents such as ethanol or acetone, and replacement of these solvents in turn with a transitional fluid such as liquid carbon dioxide by critical point drying. The carbon dioxide is finally removed while in a supercritical state, so that no gas–liquid interface is present within the sample during drying. The dry specimen is usually mounted on a specimen stub using an adhesive such as epoxy resin or electrically conductive double-sided adhesive tape, and sputter-coated with gold or gold/palladium alloy before examination in the microscope. Samples may be sectioned (with a microtome) if information about the organism's internal ultrastructure is to be exposed for imaging. If the SEM is equipped with a cold stage for cryo microscopy, cryofixation may be used and low-temperature scanning electron microscopy performed on the cryogenically fixed specimens. Cryo-fixed specimens may be cryo-fractured under vacuum in a special apparatus to reveal internal structure, sputter-coated and transferred onto the SEM cryo-stage while still frozen. Low-temperature scanning electron microscopy (LT-SEM) is also applicable to the imaging of temperature-sensitive materials such as ice and fats. Freeze-fracturing, freeze-etch or freeze-and-break is a preparation method particularly useful for examining lipid membranes and their incorporated proteins in "face on" view. The preparation method reveals the proteins embedded in the lipid bilayer.
Materials
According to Goldstein et al, "There are many ways to section the samples, including blade sawing, wire sawing, abrasive cutting, fracturing (best for more brittle materials), shearing, spark erosion, and microtomy. The surface to be prepared for microstructural analysis is polished using a graded sequence of abrasive materials. Ceramic and geological samples, like metals, may require etching to permit microstructural features to be imaged or analyzed in the SEM." However, the electron beam path from the specimen surface to ground must remain unbroken to ensure the specimen does not act like an electron mirror when sufficient charge builds up equal to that of the incident beam, which is referred to as "charging." According to Goldstein et al, "The best and simplest way to overcome charging problems is to deposit a thin metal layer on the surface of the sample." This is achieved via vacuum evaporation coating or sputter coating. For elements with atomic numbers 8 through 20, only carbon, aluminum and chromium are suitable coating materials. FIB is analogous to SEM, but with the use of positively charged ions instead of electrons. Besides imaging, FIB can be used for precision cutting, etching, or depositing conductive materials on the sample surface.
Scanning process and image formation
The SEM consists of an electron column, with the electron gun, electron lenses and vacuum pumps, plus a control console, with viewing screen and electron beam controls. The electron gun generates and accelerates electrons between 0.1 to 30 keV. Electron lenses focus the electron beam to less than 10 nm at the sample, where it interacts with the specimen to a depth of about 1 micrometre, generating signals used for point by point imaging. Deflection coils sweep the electron beam across the specimen in a raster scan fashion, with the backscattered and secondary electrons collected by the electron detector, forming an image that can be processed and displayed digitally. Sharpness and feature visibility are controlled by electron probe size, electron probe current, electron probe convergence angle and electron beam accelerating voltage. According to Goldstein et al, "Most older SEMs use tungsten or LaB 6 thermionic emitters, but, increasingly, new microscopes are equipped with cold, thermal, or Schottly field emission sources because these provide enhanced performance, reliability, and lifetime." crystal. The crystals are captured, stored, and sputter-coated with platinum at cryogenic temperatures for imaging.
Magnification
Magnification in an SEM can be controlled over a range of about 6 orders of magnitude from about 10 to 3,000,000 times. Changing magnification is achieved by adjusting the length of the scan on the specimen, with higher magnification requiring a smaller sample area, or pixel.
Detection of secondary electrons
The most common imaging mode collects low-energy ( File:Surface of a kidney stone.jpg|Surface of a kidney stone File:Surface of a kidney stone Re-colorized SEM Image.jpg|The same after re-processing of the color from the estimated topography File:Discoaster-side-diag-alt hg.jpg|SEM image of a diagenetically altered discoaster File:Discoaster-side-diag-alt Re-colorized SEM Image.jpg|The same image after similar colorization
SEM image coloring
Very often, published SEM images are artificially colored. This may be done for aesthetic effect, to clarify structure or to add a realistic appearance to the sample and generally does not add information about the specimen. Coloring may be performed manually with photo-editing software, or semi-automatically with dedicated software using feature-detection or object-oriented segmentation. Alternately, when additional information from other detectors like EDX, EBSD, ECCI or cathodoluminescence is available, it can be merged as color channel(s) to provide rich material information in a single, high-resolution image. File:Cobaea scandens1-4.jpg|SEM image of Cobaea scandens pollen File:Cobaea scandens colorized SEM image.jpg| The same after semi-automatic coloring. Arbitrary colors help identifying the various elements of the structure. File:Tradescantia tolmukakarvad ja õietolm.JPG |Colored SEM image of Tradescantia pollen and stamens File:Gold on arsenopyrite SEM image.png |Colored SEM image of native gold and arsenopyrite crystal intergrowth
Color built using multiple electron detectors
In some configurations more information is gathered per pixel, often by the use of multiple detectors. As a common example, secondary electron and backscattered electron detectors are superimposed and a color is assigned to each of the images captured by each detector, with a result of a combined color image where colors are related to the density of the components. This method is known as density-dependent color SEM (DDC-SEM). Micrographs produced by DDC-SEM retain topographical information, which is better captured by the secondary electrons detector and combine it to the information about density, obtained by the backscattered electron detector. File:DDC-SEM of calcified particle in cardiac tissue - BW - 1.jpg|DDC-SEM of calcified particle in cardiac tissue - Signal 1: SE|alt=DDC-SEM of calcified particle in cardiac tissue - Signal 1 : SE File:DDC-SEM_of_calcified_particle_in_cardiac_tissue_-_BW_-_2.jpg| Signal 2: BSE|alt=Signal 2 : BSE File:DDC-SEM of calcified particle in cardiac tissue - orange.jpg|Colorized image obtained from the two previous. Density-dependent color scanning electron micrograph SEM (DDC-SEM) of cardiovascular calcification, showing in orange a calcium phosphate spherical particle (denser material) and, in green, the extracellular matrix (less dense material) File:Cardiovascular calcification - Sergio Bertazzo.tif|Same work with a larger view, part of a study on human cardiovascular tissue calcification
Analytical signals based on generated photons
Measurement of the energy of photons emitted from the specimen is a common method to get analytical capabilities. Examples are the energy-dispersive X-ray spectroscopy (EDS) detectors used in elemental analysis and cathodoluminescence microscope (CL) systems that analyze the intensity and spectrum of electron-induced luminescence in (for example) geological specimens. In SEM systems using these detectors it is common to color code these extra signals and superimpose them in a single color image, so that differences in the distribution of the various components of the specimen can be seen clearly and compared. Optionally, the standard secondary electron image can be merged with the one or more compositional channels, so that the specimen's structure and composition can be compared. Such images can be made while maintaining the full integrity of the original signal data, which is not modified in any way.
Time resolution in SEM
There are various options for performing time-resolved measurements with an SEM. For slower time scales, it can be sufficient to scan over a sample multiple times and deduce the dynamics from this serie of images. If higher time resolution is required, a fast detector can be used. In the following example, five modes of an AFM cantilever were excited at frequencies of 32.2 kHz, 196 kHz, 522 kHz, 49 kHz and 231 kHz, and a fast detector (a Microchannel plate detector) recorded the time at which a secondary electron was detected relative to the excitation. The data was then sorted to create 100 frames for each of the shown modes.. For even faster phenomena on the picosecond timescale, pulsed electron sources are used and dynamics can be observed in a pump-probe scheme. The electron pulses can either be created with beam blankers or a laser driven photocathode.
3D in SEM
Unlike SPMs, SEMs do not naturally produce 3D images. However 3D data can be obtained using an SEM by methods such as Stereoscopy.
3D SEM reconstruction from a stereo pair
photogrammetry is the most metrologically accurate method to bring the third dimension to SEM images. Contrary to photometric methods (next paragraph), photogrammetry calculates absolute heights using triangulation methods. The drawbacks are that it works only if there is a minimum texture, and it requires two images to be acquired from two different angles, which implies the use of a tilt stage. (Photogrammetry is a software operation that calculates the shift (or "disparity") for each pixel, between the left image and the right image of the same pair. Such disparity reflects the local height). File:SEM Stereo Pair.jpg|An SEM stereo pair of microfossils of less than 1 mm in size (Ostracoda) produced by tilting along the longitudinal axis File:SEM Stereo pair of micro-fossil (Juxilyocypris schwarzbachi Ostracoda).gif|From this pair of SEM images, the third dimension has been reconstructed by photogrammetry (using MountainsSEM software, see next image); then a series of 3D representations with different angles have been made and assembled into a GIF file to produce this animation. File:3D surface reconstruction from 2 scanning electron microscope images.gif|3D surface reconstruction of a (Ra = 3 μm) roughness calibration sample (as used to calibrate profilometers), from 2 scanning electron microscope images tilted by 15° (top left). The calculation of the 3D model (bottom right) takes about 1.5 second and the error on the Ra roughness value calculated is less than 0.5%.
Photometric 3D SEM reconstruction from a four-quadrant detector by "shape from shading"
This method typically uses a four-quadrant BSE detector (alternatively for one manufacturer, a 3-segment detector). The microscope produces four images of the same specimen at the same time, so no tilt of the sample is required. The method gives metrological 3D dimensions as far as the slope of the specimen remains reasonable. Most SEM manufacturers now (2018) offer such a built-in or optional four-quadrant BSE detector, together with proprietary software to calculate a 3D image in real time. Other approaches use more sophisticated (and sometimes GPU-intensive) methods like the optimal estimation algorithm and offer much better results at the cost of high demands on computing power. In all instances, this approach works by integration of the slope, so vertical slopes and overhangs are ignored; for instance, if an entire sphere lies on a flat, little more than the upper hemisphere is seen emerging above the flat, resulting in wrong altitude of the sphere apex. The prominence of this effect depends on the angle of the BSE detectors with respect to the sample, but these detectors are usually situated around (and close to) the electron beam, so this effect is very common.
Photometric 3D rendering from a single SEM image
This method requires an SEM image obtained in oblique low angle lighting. The grey-level is then interpreted as the slope, and the slope integrated to restore the specimen topography. This method is interesting for visual enhancement and the detection of the shape and position of objects; however the vertical heights cannot usually be calibrated, contrary to other methods such as photogrammetry. File:FLY EYE.jpg|SEM image of a house fly compound eye surface at 450× magnification File:Fly eye detail.jpg|Detail of the previous image File:Fly Eye 3D SEM Image with form.jpg|SEM 3D reconstruction from the previous using shape from shading algorithms File:Fly Eye 3D SEM Image without form.jpg|Same as the previous, but with lighting homogenized before applying the shape from shading algorithms
Other types of 3D SEM reconstruction
Inverse reconstruction using electron-material interactive models Multi-Resolution reconstruction using single 2D File: High-quality 3D imaging may be an ultimate solution for revealing the complexities of any porous media, but acquiring them is costly and time-consuming. High-quality 2D SEM images, on the other hand, are widely available. Recently , a novel three-step, multiscale, multiresolution reconstruction method is presented that directly uses 2D images in order to develop 3D models. This method, based on a Shannon Entropy and conditional simulation, can be used for most of the available stationary materials and can build various stochastic 3D models just using a few thin sections. Ion-abrasion SEM (IA-SEM) is a method of nanoscale 3D imaging that uses a focused beam of gallium to repeatedly abrade the specimen surface 20 nanometres at a time. Each exposed surface is then scanned to compile a 3D image.
Applications of 3D SEM
One possible application is measuring the roughness of ice crystals. This method can combine variable-pressure environmental SEM and the 3D capabilities of the SEM to measure roughness on individual ice crystal facets, convert it into a computer model and run further statistical analysis on the model. Other measurements include fractal dimension, examining fracture surface of metals, characterization of materials, corrosion measurement, and dimensional measurements at the nano scale (step height, volume, angle, flatness, bearing ratio, coplanarity, etc.). SEM is also used by art conservationists to discern threats to paintings' surface stability due to aging, such as the formations of complexes of zinc ions with fatty acids. Forensic scientists use SEM to detect art forgeries.
Gallery of SEM images
The following are examples of images taken using an SEM. File:Soybean cyst nematode and egg SEM.jpg|Colored SEM image of soybean cyst nematode and egg. The artificial coloring makes the image easier for non-specialists to view and understand the structures and surfaces revealed in micrographs. File:Krilleyekils.jpg|Compound eye of Antarctic krill Euphausia superba. Arthropod eyes are a common subject in SEM micrographs due to the depth of focus that an SEM image can capture. Colored picture. File:Antarctic krill ommatidia.jpg|Ommatidia of Antarctic krill eye, a higher magnification of the krill's eye. SEMs cover a range from light microscopy up to the magnifications available with a TEM. Colored picture. File:SEM blood cells.jpg|SEM image of normal circulating human blood. This is an older and noisy micrograph of a common subject for SEM micrographs: red blood cells. File:HederelloidSEM.jpg|SEM image of a hederelloid from the Devonian of Michigan (largest tube diameter is 0.75 mm). The SEM is used extensively for capturing detailed images of micro and macro fossils. File:BSEGlassInclusionSb.jpg|Backscattered electron (BSE) image of an antimony-rich region in a fragment of ancient glass. Museums use SEMs for studying valuable artifacts in a nondestructive manner. File:SEGlassCorrosion.jpg|SEM image of the corrosion layer on the surface of an ancient glass fragment; note the laminar structure of the corrosion layer. File:Photoresist SEM micrograph.JPG|SEM image of a photoresist layer used in semiconductor manufacturing taken on a field emission SEM. These SEMs are important in the semiconductor industry for their high-resolution capabilities. File:Surface of a kidney stone.jpg|SEM image of the surface of a kidney stone showing tetragonal crystals of Weddellite (calcium oxalate dihydrate) emerging from the amorphous central part of the stone. Horizontal length of the picture represents 0.5 mm of the figured original. File:LightLTSEM.jpg|Two images of the same depth hoar snow crystal, viewed through a light microscope (left) and as an SEM image (right). Note how the SEM image allows for clear perception of the fine structure details which are hard to fully make out in the light microscope image. File:Onion flake. Cells. SEM-BSE.jpg|Epidermal cells from the inner surface of an onion flake. Beneath the shagreen-like cell walls one can see nuclei and small organelles floating in the cytoplasm. This BSE-image of a lanthanoid-stained sample was taken without prior fixation, dehydration, or sputtering. File:SEM-stomata-UIowa.tif|SEM image of stomata on the lower surface of a leaf
See also
References
External links
General HowStuffWorks – How Scanning Electron Microscopes Work Learn to use an SEM – An online learning environment for people wanting to use an SEM. Provided by Microscopy Australia Virtual SEM – sparkler an interactive simulation of a scanning electron microscope (SEM) Multichannel color SEM imaging and with BSE Video on the scanning electron microscope, Karlsruhe University of Applied Sciences Animations and explanations on various types of microscopes including electron microscopes (Université Paris Sud) History Environmental Scanning Electron Microscope (ESEM) history Images Rippel Electron Microscope Facility Many dozens of (mostly biological) SEM images from Dartmouth College. Lanthanoid staining SEM images from Research Institute of Eye Diseases, Moscow.