Station Outlets (US) or Terminal Units (ISO, CSA) consist of an outlet port with color-coded faceplate attached to a medical gas supply line, and primary and secondary check valves which open and close automatically upon use and disengagement from the system. In most jurisdictions, station outlets are required to be equipped with safety systems that prevent cross-connection errors, such as connecting a medical air flowmeter to an oxygen port.

Hospitals mainly use DISS connections, though a wide variety of quick-connect adapters, based on male strikers and female outlet ports, are still in use today. These latter types are easier to connect, but are more prone to leaks and failure.

Station Outlets (US) or Terminal Units (ISO, CSA) consist of an outlet port with color-coded faceplate attached to a medical gas supply line, and primary and secondary check valves which open and close automatically upon use and disengagement from the system. In most jurisdictions, station outlets are required to be equipped with safety systems that prevent cross-connection errors, such as connecting a medical air flowmeter to an oxygen port.

Hospitals mainly use DISS connections, though a wide variety of quick-connect adapters, based on male strikers and female outlet ports, are still in use today. These latter types are easier to connect, but are more prone to leaks and failure.

Chemetron/NCG

The Chemetron/NCG system resembles a simple electrical plug in appearance. Although it is very easy to connect and disconnect, this advantage poses a risk should it be dislodged accidentally. Additionally, its straight valve does not allow for the connection of flowmeters, necessitating the purchase and use of additional equipment.

Hansen Schrader

This unusual system consists of steel strikers and ports. The ports have a spring-loaded collar, which will be pulled back to expose a series of balls located on the inner lip of the port. The positioning of these balls is specific to each gas.

At this point, the male striker may be inserted. It is possible this system never caught on due to its lack of colour-coding and labels, along with the difficulty of engaging the striker.

Oxequip Medstar/Canadian Liquid Air (CLA)

These consist of steel strikers ("bayonets") with holes for inlet ports. Like keys, the strikers have grooves and ridges in variable positions. They lack colour-coding and are easily disengaged.

See also

Medical gas supply

References

Branson, Richard D., et al. 'Respiratory Care Equipment, 2nd Edition'. Cont. Banner et al. pp. 35–36.

(Philadelphia: Lippincott Williams & Wilkins, 1999). Cairo, J.M., et Pilbeam, Susan. 'Mosby's Respiratory Care Equipment'.

Ed. Kelli Chronister et al. pp. 47. (St. Louis, Missouri: Mosby Elsevier, 2010). Wilkins, Robert L. et al. 'Egan's Fundamentals of Respiratory Care, Ninth Edition'.

Ed. Lucy Kester et al. pp. 862–865, 898-899. (St. Louis, Missouri: Mosby Elsevier, 2009).

Text derived from https://en.wikipedia.org/w/index.php?oldid=1337124240 under CC BY-SA 4.0; modifications have been made.

Provenance

snapshot enwiki-001337124240 · captured 1770480400.0 · version history

Cite this page as: encyclopegia.org, “Station Outlet”, snapshot enwiki-001337124240. https://encyclopegia.org/a/station-outlet

Sources cited in this revision

No external citation was extracted for this revision. That is a statement about the source, not a claim that none exists.