Biotransducer
≈ 5 min read
Names in other languages
- العربيةمحول الطاقة الحيوية
- EnglishBiotransducer
- فارسیمبدل زیستی
Article bodies are English and are not machine-translated. A name in another language is not a translation of the article.
A biotransducer is the recognition-transduction component of a biosensor system. It consists of two intimately coupled parts: a bio-recognition layer and a physicochemical transducer, which act together to convert a biochemical signal to an electronic or optical signal. The bio-recognition layer typically contains an enzyme or another binding protein, such as an antibody.
However, oligonucleotide sequences, sub-cellular fragments such as organelles (e.g. mitochondria) and receptor-carrying fragments (e.g. cell wall), single whole cells, small numbers of cells on synthetic scaffolds, or thin slices of animal or plant tissues may also comprise the bio-recognition layer. It gives the biosensor selectivity and specificity.
The physicochemical transducer is typically in intimate and controlled contact with the recognition layer. As a result of the presence and biochemical action of the analyte (target of interest), a physicochemical change is produced within the biorecognition layer that is measured by the physicochemical transducer, producing a signal that is proportionate to the concentration of the analyte.
The physicochemical transducer may be electrochemical, optical, electronic, gravimetric, pyroelectric, or piezoelectric. Based on the type of biotransducer, biosensors can be classified as shown to the right.
A biotransducer is the recognition-transduction component of a biosensor system. It consists of two intimately coupled parts: a bio-recognition layer and a physicochemical transducer, which act together to convert a biochemical signal to an electronic or optical signal. The bio-recognition layer typically contains an enzyme or another binding protein, such as an antibody.
However, oligonucleotide sequences, sub-cellular fragments such as organelles (e.g. mitochondria) and receptor-carrying fragments (e.g. cell wall), single whole cells, small numbers of cells on synthetic scaffolds, or thin slices of animal or plant tissues may also comprise the bio-recognition layer. It gives the biosensor selectivity and specificity.
The physicochemical transducer is typically in intimate and controlled contact with the recognition layer. As a result of the presence and biochemical action of the analyte (target of interest), a physicochemical change is produced within the biorecognition layer that is measured by the physicochemical transducer, producing a signal that is proportionate to the concentration of the analyte.
The physicochemical transducer may be electrochemical, optical, electronic, gravimetric, pyroelectric, or piezoelectric. Based on the type of biotransducer, biosensors can be classified as shown to the right.
Amperometric
Amperometric transducers detect change in current as a result of electrochemical oxidation or reduction. Typically, the bioreceptor molecule is immobilized on the working electrode (commonly gold, carbon, or platinum). The potential between the working electrode and the reference electrode (usually Ag/AgCl) is fixed at a value and then current is measured with respect to time.
The applied potential is the driving force for the electron transfer reaction. The current produced is a direct measure of the rate of electron transfer. The current reflects the reaction occurring between the bioreceptor molecule and analyte and is limited by the mass transport rate of the analyte to the electrode.
Potentiometric
Potentiometric sensors measure a potential or charge accumulation of an electrochemical cell. The transducer typically comprises an ion selective electrode (ISE) and a reference electrode. The ISE features a membrane that selectively interacts with the charged ion of interest, causing the accumulation of a charge potential compared to the reference electrode.
The reference electrode provides a constant half-cell potential that is unaffected by analyte concentration. A high impedance voltmeter is used to measure the electromotive force or potential between the two electrodes when zero or no significant current flows between them.
The potentiometric response is governed by the Nernst equation in that the potential is proportional to the logarithm of the concentration of the analyte.
Impedance
Electrochemical impedance spectroscopy (EIS) involves measuring resistive and capacitive changes caused by a biorecognition event. Typically, a small amplitude sinusoidal electrical stimulus is applied, causing current to flow through the biosensor. The frequency is varied over a range to obtain the impedance spectrum.
The resistive and capacitive components of impedance are determined from in phase and out of phase current responses. Typically, a conventional three-electrode system is made specific to the analyte by immobilizing a biorecognition element to the surface. A voltage is applied and the current is measured.
The interfacial impedance between the electrode and solution changes as a result of the analyte binding. An impedance analyzer can be used to control and apply the stimulus as well as measure the impedance changes.
Conductometry
Conductometric sensing involves measuring the change in conductive properties of the sample solution or a medium. The reaction between the biomolecule and analyte changes the ionic species concentration, leading to a change in the solution electrical conductivity or current flow.
Two metal electrodes are separated at a certain distance and an AC potential is applied across the electrodes, causing a current flow between the electrodes. During a biorecognition event the ionic composition changes, using an ohmmeter the change in conductance can be measured. \Delta V = \omega PAr\Delta T\left(1+\omega^2\tau_E^2\right)^{-1/2} \tau_E=rC where V = Voltage, ω = angular frequency of the modulated incident, P = pyroelectric coefficient, L = film thickness, ε = film dielectric constant, A = area of film, r = resistance of the film, C = capacitance of the film, τE = electrical time constant of the detector output.